Spec to bluesky
When transitioning between Spec and Bluesky software for controlling experiments, adjustments are needed for interfacing with Keithley meters. These meters use a GPIB cable with Spec and an RS232 port with Bluesky.
Switching from Spec to Bluesky
Preparation
Before switching from Spec to Bluesky: - Take a photo of the GPIB connections to ensure you can reconnect them correctly when switching back.
Steps for Switching
- Disconnect the GPIB cable from each Keithley meter.
- Connect the RS232 cable (Moxa) to the corresponding ports:
- Keithley 1: port 1
- Keithley 2: port 3
- Keithley 3: port 5
- Keithley 4: port 7
Configuration for Each Keithley Model
Keithley 6514
- Turn off GPIB communication.
- Change the communication protocol to RS232.
- Set the baud rate to 11920.
- Change the terminator setting from CR to CRLF.
- Restart the device.
Keithley 6517B
- Change the communication protocol from GPIB to RS232.
- Set the baud rate to 19200.
- Change the terminator setting from CR to CRLF.
- Restart the device.
Switching from Bluesky to Spec
To revert to using Spec after operating with Bluesky:
- Disconnect the RS232 cables from all Keithley meters.
- Reconnect the GPIB cables according to the positions noted in the photos taken during the initial switch.
- Configure each Keithley meter to use GPIB communication by reversing the changes made when switching to Bluesky.